FDADecember 18, 2015device

arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

What to do

FDA enforcement status: Terminated

Brands named

arkray factoryarkray