FDANovember 3, 2015device

VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916, Software Product Code 6802864, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

What to do

FDA enforcement status: Terminated

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

10758750009916

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916, Software Product Code 6802864, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of... — Recall Details · AllClear