FDANovember 3, 2015device

VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Unique Device Identifier Number 10758750001644, Software Version 2.8 and below, IVD. Product Usage: The VITROS 5,1 FS Che...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

What to do

FDA enforcement status: Terminated

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

1075875000113210758750001644

Recall history