C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
What to do
FDA enforcement status: Terminated
Brands named
integra lifesciencesintegra
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDACodman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane catheter for diverting cerebrospinal fluid (CSF) from the ventricles of the brain through a series of drainage holes n...2026-04-10
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EO2026-03-27
- FDACytal Burn Matrix 10x15 cm. Product ID: BMM1015.2026-03-19
- FDACytal Wound Matrix 2-Layer 5x5 cm. Product ID: WSM0505.2026-03-19
- FDACytal Wound Matrix 2-Layer 7x10 cm. Product ID: WSM0710.2026-03-19
- FDACytal Burn Matrix 7x10 cm. Product ID: BMM0710.2026-03-19
- FDACytal Wound Matrix 2-Layer 10x15 cm. Product ID: WSM1015.2026-03-19
- FDACUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.2026-03-06
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →