FDASeptember 4, 2014device

Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is e...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.

What to do

FDA enforcement status: Terminated

Brands named

midmark corp dba progenymidmarkmidmark corp