FDANovember 18, 2019device

Habib EUS RFA, 20mm electrode active tip, Material Number/UPN: 6700 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.

What to do

FDA enforcement status: Terminated

Brands named

boston scientificboston

UPCs

5060271140596