FDADecember 15, 2015device

GE Healthcare, Aisys CS2,1011-9050-000; Avance Amingo,1009-9002-000, & Avance CS2, 1009-9050-000. The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intend...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

What to do

FDA enforcement status: Terminated

Brands named

ge medical systems

UPCs

101190500001009900200010099050000

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.