FDANovember 20, 2024device

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

What to do

FDA enforcement status: Ongoing

Brands named

nova biomedicalnova

UPCs

10385480636858

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.