FDADecember 4, 2018device

DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skel...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.

What to do

FDA enforcement status: Terminated

Brands named

agfa gevaert n vagfaagfa gevaert