FDAOctober 25, 2024device

Galaxy System (GAL-001)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

What to do

FDA enforcement status: Ongoing

Brands named

noah medicalnoah