FDADecember 15, 2015device

Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.

What to do

FDA enforcement status: Terminated

Brands named

mindray ds usa inc dba mindray north americamindraymindray ds