FDADecember 23, 2021device

da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.

What to do

FDA enforcement status: Ongoing

Brands named

intuitive surgicalintuitive

UPCs

00886874112588

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.