FDANovember 6, 2023device

Curaplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

What to do

FDA enforcement status: Ongoing

Brands named

sarnova hcsarnova