FDANovember 10, 2016device

Barbiturates Intercept Micro-Plate EIA (enzyme immunoassay, barbiturate) Product is intended for use by clinical laboratories in the qualitative determination of barbiturates in oral fluids.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Shipping error. Cocaine metabolite Microplate distributed instead of Barbiturate Microplate due to incorrect selection of Microplate during manufacturing.

What to do

FDA enforcement status: Terminated

Brands named

orasure technologiesorasure