FDANovember 24, 2021device

Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

Recall history