FDANovember 18, 2019device

RayStation stand-alone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0 - Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit s...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.

What to do

FDA enforcement status: Terminated

Brands named

raysearch laboratories abraysearchraysearch laboratories