FDANovember 16, 2023device

PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.

What to do

FDA enforcement status: Ongoing

Brands named

becton dickinsonbectonbecton dickinson

UPCs

30382903065470