The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinders reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
What to do
FDA enforcement status: Terminated
Brands named
horiba instrumentshoriba
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAeTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-1582026-03-02
- FDAeTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-1592026-03-02
- FDAeTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-1562026-03-02
- FDAeTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-1512026-03-02
- FDAeTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-1572026-03-02
- FDAeTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-1522026-03-02
- FDAeTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-1502026-03-02
- FDAeTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-1492026-03-02
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →