FDANovember 18, 2019device

STA SYSTEM CONROL N + P (ref. 00678)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

What to do

FDA enforcement status: Terminated

Brands named

diagnostica stagodiagnostica