Device Description: FACSLyric 2L6C Instrument US, Catalog Number 662876, UDI 00382906628763 The K170974 clearance document describes the device as follows: The BD FACSLyric" flow cytometer (3-1, 4-2, 4-2-2, and 4-3-3 optical configurations) systems consist of a flow cytometer, sheath tank, waste ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient m
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCVeseacky Pajama Sets Recalled Due to Risk of Serious Injury Death from Burn Hazard; Violate Mandatory Standards for Children's Sleepwear2026-06-18
- FDABD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405609.2026-04-27
- FDABD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405637, 405707.2026-04-27
- FDABD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405671.2026-04-27
- FDABD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405658.2026-04-27
- FDABD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 405723, 406704.2026-04-27
- FDABD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.2026-04-27
- FDABD¿Spinal Tray with BD¿Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405632.2026-04-27
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.