HeartWare HVAD Pump Kit, REF MCS1705PU
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
What to do
FDA enforcement status: Ongoing
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDA1104CA-CLIN - HVAD (HeartWare Ventricular Assist Device) PUMP, Model Number 1104CA-CLIN2023-10-31
- FDAPUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT, Model Number 11042023-10-31
- FDAHVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT/IMPLANT KIT OUS, Model Number 12052023-10-31
- FDAPUMP 1103 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT, Model Number 11032023-10-31
- FDAHVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 11002023-10-31
- FDAHVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 11022023-10-31
- FDASTERILE HVAD (HeartWare Ventricular Assist Device) PUMP (US), Model Number 11012023-10-31
- FDAPUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT EN JA, Model Number 1104JP2023-10-31
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →