FDANovember 20, 2023device

Calibration Serum Level 3 CAL2351

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

What to do

FDA enforcement status: Ongoing

Brands named

randox laboratoriesrandox

UPCs

05055273200966