FDAJanuary 9, 2019device

NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reco...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.

What to do

FDA enforcement status: Terminated

Brands named

neusoft medical systems imp exneusoftneusoft medical