FDASeptember 28, 2016device

ANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF 06597103 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

What to do

FDA enforcement status: Terminated

Brands named

angiodynamics

Recall history