FDAJanuary 21, 2016device
Endobon-Xenograft Granules, Zimmer Biomet, Synthetic Bone Grafting Material, used in dental and/or surgical procedures. Endobon Xenograft Granules, Bone Grafting Material, Synthetic are used in the following dental and/or surgical procedures: " Alveolar ridge augmentation/reconstruction, " Sinus ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Cytotoxicity tests were not passed at 36 months. Use of Endobon-Xenograft Granules past 18 months of its specified shelf life may result in patient illness/injury.
What to do
FDA enforcement status: Terminated
Brands named
biomet france sarlbiometbiomet france
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV li...2026-05-21
- FDABIOMET ZipTight, Acute AC Joint Implant, Single Ziploop , Model/Catalog Number: 904834; Fixation System for AC Joint Repair2025-09-23
- FDAJuggerknot Mini Soft Anchors, Model/Catalog Number: 9120802025-09-09
- FDABD PhaSeal Injector Luer (N30C), REF: 515001 BD PhaSeal Injector Luer Lock (N35), REF: 515003 BD PhaSeal Injector Luer Lock (N35C), REF: 515004 BD PhaSeal Injector Lu63er Lock (N35C Multi), REF: 515005 BD PhaSeal Protector (P14), REF: 515100 BD PhaSeal Protector (P21), REF: 515102 BD PhaSeal Prot...2025-06-16
- FDABD Texium: 3 mL, REF: MY8003-0006; 5 mL, REF: MY8005-0006; 10 mL, REF: MY8010-0006; 20 mL, REF: MY8020-0006; 30 mL, REF: MY8030-0006; 60 mL, REF: MY8060-00062025-05-12
- FDAAlaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Li...2025-05-12
- FDABD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY80602025-05-12
- FDABD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-05002025-05-12