FDADecember 20, 2018device

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing usi...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.

What to do

FDA enforcement status: Terminated

Brands named

abbott gmbh co kgabbottabbott gmbh