GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typica...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.

What to do

FDA enforcement status: Terminated