FDADecember 15, 2020device
Pinnacle Cups, Part numbers 121701048 121701050 121701052 121701054 121701056 121701058 121701060 121701062 121711048 121711052 121712050 121712052 121712056 121722048 121722050 121722052 121722054 121722056 121722058 121722060 121722062 121731048 121731050 121731052 121731054 121731056 121731058...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.
What to do
FDA enforcement status: Terminated
Brands named
depuy orthopaedicsdepuy
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 5 CEM. Part Number: 150450205.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 3 CEM. Part Number: 150450203.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 5 CEM. Part Number: 150450105.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 4 CEM. Part Number: 150450104.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 7 CEM. Part Number: 150450207.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 8 CEM. Part Number: 150450208.2026-04-03