Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What to do

FDA enforcement status: Ongoing

Brands named

american contract systemsamericanamerican contract

UPCs

0019107216424000191072183255

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL — Recall Details · AllClear