Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE IMPLANT, kit number AGDI40J; c) LAB DEVICE IMPLANT, kit number AGDI40K; d) MINOR VASCULAR, kit number AGMV26O; e) OPEN HEART ACCESSORY, kit number AGOH98Q; f) OPEN HEART ACCE...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What to do

FDA enforcement status: Ongoing

Brands named

american contract systemsamericanamerican contract

UPCs

00191072475780019107215466800191072188168001910721797770019107215941900191072169754001910721797840019107217979100191072137586001910721798070019107218269200191072191472

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE IMPLANT, kit number AGDI40J; c) LAB DEVICE IMPLANT, kit number AGDI40K; d) MINOR VASCULAR, kit number AGMV26O; e) OPEN HEART ACCESSORY, kit number AGOH98Q; f) OPEN HEART ACCE... — Recall Details · AllClear