FDAFebruary 4, 2016device

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

What to do

FDA enforcement status: Terminated

Brands named

draeger medicaldraeger