BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184; UDI -008427680131024718420190221; BC Thrombin Reagent; Test, Thrombin Time Product Usage: For in vitro diagnostic use. Reagent for the quantitative determination of the thromb...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The BC Thrombin Reagent kit lot 46751 (contains Thrombin Reagent lot 517468) and kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged Thrombin Time (TT) results for expected normal samples, and may recover above the upper limit of normal (< 21 sec.) as stated within the IFU (Instruction For Use).

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare