FDAJanuary 14, 2016device

9131-001 Defibrillation Electrodes, Lot No. 141125-02 Used in Powerheart G3 9300A, 9300E, 9300P, 9390A, and 9390E AEDs Cardiac Science 9131 Defibrillation Electrodes are single use and intended to be used in conjunction with Cardiac Science automatic external defibrillators (AED) to monitor and d...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Defibrillation electrodes may increase electrical impedance over time. If impedance becomes too high, the AED will fail the electrode self-test and it will not be Rescue Ready, the AED STATUS INDICATOR will be RED and the AED will beep. This is normal behavior when the impedance limit is exceeded. The high impedance electrodes require replacement.

What to do

FDA enforcement status: Terminated

Brands named

cardiac sciencecardiac