Impella RP with SmartAssist; Product Number: 0046-0035;
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAbiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834.2026-05-22
- FDAAbiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.2026-05-22
- FDAAbiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000435. Introducer Product Code: 1000435.2026-05-22
- FDAAbiomed 14 Fr x 13 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000434. Introducer Product Code: 1000434.2026-05-22
- FDAImpella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.2026-05-18
- FDAAutomated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packa...2026-05-14
- FDAAutomated Impella Controller (AIC); Product Code: 0042-0000-US;2026-04-20
- FDAPurge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge Cassette, Packaged; Product C...2026-02-18