FDAJanuary 11, 2016device

Gender PFJ FEMORAL COMP, prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer various sizes " Osteoarthritis, traumatic arthritis, polyarthritis, and/or severe chondrocalcinosis of the patellofemoral joint. The salvage of previously failed su...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

LDPE bag containing the implant adheres to the highly polished implant surface.

What to do

FDA enforcement status: Terminated

Brands named

zimmer biometzimmer

UPCs

00592601101005926011020059260120100592601202005926013010059260130200592601401005926014020059260150100592601502