FDAJanuary 31, 2022device

Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.