FDAJanuary 17, 2019device

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

What to do

FDA enforcement status: Terminated

Brands named

medtronic inc cardiac rhythm and heart failure crhfmedtronicmedtronic inc

UPCs

00643169121973064316916697400643169166981006431691669980064316916700100643169167018006431691670250064316916704900643169167063006431691670700064316916708700643169167094