FDANovember 28, 2023device

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

What to do

FDA enforcement status: Ongoing

Brands named

xtant medicalxtant

UPCs

0081007656400515010085050