FDAFebruary 22, 2022device

Illumina NextSeq 550Dx, REF: 20005715, CE, IVD

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

What to do

FDA enforcement status: Completed

Brands named

illumina

UPCs

00816270020125