FDADecember 16, 2020device

REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction

What to do

FDA enforcement status: Ongoing

Brands named

medtronic neurosurgerymedtronic