FDAJanuary 28, 2022device

Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.

What to do

FDA enforcement status: Terminated

Brands named

b braun medical

UPCs

04046963686345040469642856080404696428561504046963686338

Recall history