FDAOctober 30, 2020device

Philips Respironics V60 Ventilator Part Number 1053614 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.

What to do

FDA enforcement status: Terminated

Brands named

respironics californiarespironics