FDADecember 18, 2024device

GE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2) H2401AK; 3) H2402AC; 4) H3100AH; 5) H3100XB; System, Tomography, Computed, Emission

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

What to do

FDA enforcement status: Ongoing

Brands named

ge medical systems israel functional imaging

UPCs

5376204705428313162548083119512125831512043418315361190183136989873831147013608316299027685150299222840682140904840682121194082426210029