FDAJanuary 20, 2022device

BenchMark ULTRA and DISCOVERY ULTRA Instruments

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.

What to do

FDA enforcement status: Ongoing

Brands named

ventana medical systemsventanaventana medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.