FDADecember 18, 2024device

GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Emission

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

What to do

FDA enforcement status: Ongoing

Brands named

ge medical systems israel functional imaging

UPCs

82426110010082426040045082426110007082426190025082426220008082426140021082426230023082426120017082426140026082426150025082426130031082426160036

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.