FDAMarch 1, 2022device
AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A, K09-12917, K09-13501AP, K10-05457AP, K10-05635, K10-05868P, K10-05872P, K12-08303, K12-10346, K12-10375, K12-10724AP, K12-10962, K12-11413, MAP150/F, MAP152/B, MAP...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.
What to do
FDA enforcement status: Ongoing
Brands named
merit medical systemsmeritmerit medical
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAGSMS Incorporated, NIACIN EXTENDED-RELEASE TABLETS, USP, 1,000 MG, 90 tablets, Rx only, Manufactured by Kremers Urban Pharmaceuticals Inc., a subsidiary of Lannett, Inc., Seymour, IN 47274, Packaged by GSMS Incorporated, Camarillo, CA 93012. NDC 51407-268-90.2026-05-29
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18
- CPSCGenerac Power Systems Expands Recall of Generac and DR Power Electric Start Pressure Washers Due to Risk of Serious Injury or Death from Carbon Monoxide Hazard; Includes Additional Models2026-05-14
- FDANicotine Transdermal System Patch, 14 mg, 14 patches per box, Aveva, Distributed by: RUGBY LABORATORIES, Indianapolis, IN 46268, www.majorpharmaceuticals.com, NDC 0536-5895-88.2026-05-14
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAMerlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software2026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →