FDADecember 8, 2017device

ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL....

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.

What to do

FDA enforcement status: Terminated

Brands named

abbott laboratoriesabbott

Recall history