FDADecember 11, 2019device
Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M123...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.
What to do
FDA enforcement status: Terminated
Brands named
lumitex
Recall history
No related federal recalls on record for this brand yet.