FDAOctober 22, 2019device

BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.

What to do

FDA enforcement status: Terminated

Brands named

biomerieux