FDAOctober 2, 2018device

AF550x5ML/H 2ML 60LO 6" CATH, 4.5" IN, Product code MT5060XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

What to do

FDA enforcement status: Terminated

Brands named

arrow